BINDER incubator - BF series BF 720 Liter
Overview
Incubator with forced convection
The BINDER incubator BF series is the highest-class for all gentle incubation tasks, particularly under a full load and large batch throughputs. It is extremely homogeneous and has quick recovery times.
Performance Characteristics
Electronically controlled APT.lineâ„¢ preheating chamber assuring temperature accuracy and reproducible results
Temperature range from 5 ° C above room temperature to 100 ° C
Digital temperature setting with an accuracy of a tenth of a degree
MS controller with several timer functions
Controller time functions:
0 to 99 h 59 min delayed ON
0 to 99 h 59 min delayed OFF
0 to 99 h 59 min temperature-dependent
delayed OFF
Adjustable fan speed
Adjustable ventilation by means of front ventilation flap slide and rear exhaust Ø 50 mm
Inner glass door
Independent temperature safety device class 3.1 ( DIN 12880) with optical alarm
RS 422 interface for use with APT-COMâ„¢ DataControlSystem communication software or switch over to printer output with RS 232 / RS 422 interface converter
Adjustable intervals for printer
Units up to 115 liters are stackable
2 chrome-plated racks included
BINDER test confirmation
Specifications :
BF 720
Exterior dimensions
Width ( mm) 1235
Height ( incl. casters) ( mm) 1530
Depth ( mm) 865
Plus door handle and connection ( mm) 90
Wall clearance, rear ( mm) 100
Wall clearance, side ( mm) 160
Exhaust duct ( outer Ø mm) 52
Steam space volume ( l) 808
Number of doors ( ea.) 2
Interior dimensions
Width ( mm) 1000
Height ( mm) 1200
Depth ( mm) 600
Interior volume ( l) 720
Racks ( number standard/ max.) 2 / 15
Load per rack ( kg) 45
Permitted total load ( kg) 120
Weight ( empty) ( kg) 180
Temperature data
Temperature range approx. 5 ° C above ambient temperature to ( ° C) 100
Temperature variation
at 37 ° C ( ± K) 0, 4
at 50 ° C ( ± K) 0, 6
Temperature fluctuation
at 37 ° C ( ± K) 0, 1
at 50 ° C ( ± K) 0, 2
Warm-up time 1)
to 37 ° C ( min.) 21
to 50 ° C ( min.) 24
Recovery time after doors were open for 30 sec. 1)
at 37 ° C ( min.) 1
at 50 ° C ( min.) 4
Ventilation at 70 ° C ( x/ h) 11
Electrical data
IP protection class acc. to EN 60529 IP 20
Voltage ( ± 10% ) 50 / 60 Hz ( V) 230
Nominal power ( kW) 1, 25
Energy consumption at 37 ° C ( W) 80
1) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 ° C and a line voltage fluctuation of ± 10% . The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.