    Biochemical Incubator

    Biochemical Incubator

    Quantity Order:

    Specification :

    60A / 150A / 250B type of raising boxes. Bake born
    Product synopsis:
    Biochemical incubator with cooling and heating temperature controlled system, two-way temperature-adjusting, is the function of plant, animal, microbial, genetic, environmental protection, medical, viruses, scientific research, teaching department indispensable equipment, widely used in laboratory experiments, the constant training experiment, environment experiments, etc. Its main features: the product adopts integrated circuit control temperature, LED digital display, clearly.
    The technical indexes:
    1. Use power: isolator 220V
    2 105W 150A refrigeration power: ( ) , 125W ( 250B)
    3 heating power: 100W ( 150A) , 150W ( 250B)
    4 control temperature range: 0-50 ° c ( digital LED with observation window) ,
    May: 0.5 degrees Celsius temperature control accuracy ±
    6 60L: 150L/ effective volume 250L /
    7. Working time: continuous
    Instructions:
    1 incubator should be placed a dry, clean and ventilated, far away from heat and sunlight work placement, smooth, prevent indoor noise vibration.
    2 before using, panel each control switch in non-chip-producing state.
    3 in the incubator loaded with experimental samples, when the experiment bottle ( or vessels) adequate distance between ( heat) , air convection cycle.
    4. Turn ON the power switch, will power " open" ( ON) position indicator light, LED digital display that shows in actual temperature.
    5 " switch" digital dial the code you need to control by the temperature, then also closed " heat" refrigeration " switch, when the temperature below the actual cases, will be heated up, When the temperature in the cabinet above the value, will automatically refrigeration cooling, incubator can automatic constant temperature control. One on the control panel, red, green, respectively, the indicator that " heat" and " cooling" working state.
    Maintenance and repair:
    1 incubator should reliable grounding.
    2. To ensure the condensing heat, condenser and effectively than 100mm distance between the wall. Box side, the top 50 clearance should be at least 30mm space.
    3 incubator in handling, repair, maintenance, should avoid violent collisions, less than 45 degrees to45degrees) .
    4 power socket plucked after 5 minutes to insert again, such as soon, compressor parts into damaged.
    5 the incubator, unfavorable make refrigeration temperature in the oven and environmental temperature difference than 25 degrees.
    6 containing volatile chemical is prohibited and explosive gas and solvent combudstible gas in box, do not use near incubator, lest spark ignites combustible spray.
    7. To prevent pollution, low temperature should be used in the working chamber wall to avoid condensation frost.
    8. Refrigeration system work, should avoid to open the door and the frequent keep temperature stability, prevent the dust, filth from entering. All
    9 refrigeration system after stop working, with soft cloth rub-up studio.