Specification :
Special Feature
Display: 5.5’ SNT screen
Patient type: Adult, Pediatric
Driven mode: Pneumatically driven electronically controlled
Mode: Semi-open, Semi-closed or closed system
Anesthetic vaporizer: Two high precise vaporizers
Power supply: 220V± 10% , 50Hz± 1Hz
110V± 10% , 60Hz± 1Hz
Iternal battery: DC12V ( for 2 hours)
Power consumption: d50W
Anesthesia Ventilator
Respiration mode: IPPV, SIMV, SIGH, MANUAL
Waveform: P-t, V-t
Tidal volume: 50ml~ 1600ml
Respiratory frequency: 1 to 65 BPM
I: E ratio: 8: 1~ 1: 8
Alarm parameters:
High pressure alarm, Low pressure alarm, O2 source failure alarm, AC power failure alarm
High Pressure alarm limit: 20~ 60cmH2O Low Pressure alarm Limit: 0~ 20cmH2O
Monitor Parameters: Tidal volume, MV, Ppeak, PEEP, respiratory frequency, I: E ratio, Airway pressure
Safety valve pressure limit: 6Kpa
Good air tightness, high reliability, and minor respiration impedance
Main Unit
Circle absorber: Compact circle absorber
Flow meter:
O2, N2O four-tube flow meter
O2 tube: 0L/ min-1L/ min~ 10L/ min.
N2O tube: 0L/ min~ 1L/ min~ 10L/ min
O2 and N2O Linkage Device: Adjust O2 and N2O in proportion ( keep O2 concentration not lower than 25% )
Oxygen flush: 25L/ min~ 75L/ min
Gas source pressure:
O2: 0.25Mpa~ 0.65MPa.
N2O: 0.25Mpa~ 0.65MPa.
Gauge: Two gauges, N2O, O2