ELE, Automatic Vicat[Apr. 13, 2011 22:25:10]
Automatic Vicat
EN 196-3, 13454-2; ASTM C187, C191
The Automatic Vicat Apparatus executes the test program
using a fully automatic test cycle. The integral LCD
display indicates test progress in real time. Firmware
in the unit enables up to five user test profiles to be
established. Integrated on-board memory will store up to
50 complete tests. On completion of the test the integral
printer automatically prints all test data including a graph
penetration with related data. As standard, the unit
includes an RS232 serial port for connection to PC which,
when used in conjunction with the supplied software,
enables users to manage test data including graphing
and report generation. The timed cycle of events is
operator-selectable and penetrations can be selectable at
intervals of 30 seconds, 1, 5, 15 or 30 minutes.
Ordering Information
EL38-2015/ 01 Automatic Vicat
Apparatus complete
with EN and ASTM Initial and Final needles, Consistency
Plunger, Windows® software and RS232 cable, 1 x EN
and ASTM Mould and Glass Plate. Suitable for continuous
use in saturated humidity, at a controlled temperature of
20° C ± 1° C. Weight 13 kg.
For 220 – 240 V AC, 50 – 60 Hz, 1 ph.
Accessories
EL38-2015/ 12 Needle Cleaning Device
EL38-2015/ 14 Printer Paper Rolls. Pack of 10.
EL38-2015/ 15 Mould Tank for testing samples under
water as per EN 196-3, for use in temperature controlled
laboratories.
Spares
EL38-2015/ 16 Consistency Plunger
EL38-2021 comprising 1 x EN Initial Set Needle, 1.13
mm dia and 1 x EN Final Set Needle, 1.13 mm dia.
EL38-2023 comprising 1 x ASTM Initial Set Needle, 1
mm dia and 1 x ASTM Final Set Needle 1 mm dia.
Moulds see EL38-2200, EL38-2300 or EL38-2660