Welcome to Indotrade.co
Sign In | Sign Up

Welcome to

MIKASE LABORATORY - PT MITRA KATIGA SEJAHTERA / SUPPLIER ALAT UJI KUALITAS UDARA, AIR, CUACA , KEBISINGAN SERTA GETARAN

http://indotrade.co/mikaselaboratory/

Priority Member Indotrade Priority
17th Year

showing 95 of 191
  • Products Catalog

    THREE AXIS ULTRASONIC ANEMOMETER, Model : HD2003 and HD2003.1, Brand : DeltaOhm

    [Nov. 4, 2010 1:47:55]
    THREE AXIS ULTRASONIC ANEMOMETER,  Model : HD2003 and HD2003.1,  Brand : DeltaOhm
    Quantity Order
    Add to Basket
    Origin
    Italy
    Recommend to your friends Put on Product Partner

    Specification

    HD2003 and HD2003.1 are three axis ultrasonic anemometers, they measure
    the speed and direction of wind, the U-V-W Cartesian components of speed, sound speed and sonic temperature.
    The HD2003 allows also to detect temperature and relative humidity of the air and barometric pressure.
    The HD2003 main features are:
    • Determination of the anemometric quantities represented in diverse measurement
    units: wind speed and direction, U-V-W Cartesian components of speed, sound speed, sonic temperature.
    • ( HD2003 Model) additional output quantities: Temperature, Relative Humidity
    and Pressure.
    • 5 analogue voltage or current outputs, with different measuring ranges.
    • RS232 and Multidrop RS485 Serial Communication interfaces.
    • Configurable output rate of digital output data string.
    • Configurable average periods 1÷ 60sec and 1÷ 60min. for all output quantities.
    • Algorithmic raw date processing and validation, assuring ± 1% precision to
    anemometric quantities.
    • Digital high frequency data acquisition mode with 50Hz data output.
    • Self-diagnosis with error checking and report.
    • Reliability and precision on whole measuring range, no additional calibration required.
    • Flexible, easy-to use operating software, confi gurable according to the user’ s
    needs through Computer interface.
    • User interface for ‘ Setup’ management and software upgrade through RS232 or
    RS485.
    • Automatic alignment to the magnetic North through built in compass.
    • No moving part, with reduced maintenance and service costs.
    • Rugged and reliable structure, suitable for continuous operation even in severe environmental conditions.
    • Low power consumption.
    • ( On request) Heaters Option: built-in heating device of sonic transducers, to prevent ice and snow formation. Assures correct measurements even in presence of sleet or snow.
    Typical applications:
    - Meteorology
    - Aviation and Navigation
    - Tunnels, Highways
    - Climatology
    - Sport and winter stations
    - Safety in yards
    - Industrial buildings
    Technical specifications
    Output quantities
    • Anemometric parameters Wind speed and direction, Sound Speed, Sonic Temperature, U-V-W Components
    • Meteorological parameters ( Model HD2003) Pressure, Temperature, Relative Humidity
    • Heading Compass with magnetic Azimuth
    • Moving Averages 1÷ 60 sec./ 1 ÷ 60 min.
    • Output rate 1÷ 50Hz ( RS232 or RS485)
    Wind Speed
    • Measuring unit m/ s, cm/ s, km/ h, knots, mph
    • Range 0÷ 60 m/ s ( 216 km/ h)
    • Resolution 0.01 m/ s
    • Accuracy ± 1% of reading
    Wind Direction
    • Range Azimuth: 0÷ 360° Elevation: ± 60°
    • Resolution 0. 1°
    • Accuracy ± 1°
    Sound speed
    • Range 300 ÷ 380 m/ s
    • Resolution 0.01 m/ s
    • Accuracy ± 1% of reading
    Sonic Temperature
    • Range -40 + 60° C
    • Resolution 0.1 ° C
    • Accuracy ± 1° C
    Compass
    • Range 0 ÷ 360°
    • Resolution 0.1 °
    • Accuracy ± 1°
    Digital Outputs
    • Communications RS-232 full duplex, Multidrop RS-485 half duplex
    • Baud Rate 9600 ÷ 115200 bit/ sec.
    • Output Rate Normal functioning mode: 1 ÷ 3600 sec
    Digital high frequency: 1/ 50 sec
    • Measured data Digital string of anemometric quantities and compass ( Model HD2003) Pressure, temperature, relative
    humidity
    Relative Humidity
    Capacitive sensor
    Analog output ( 0 ÷ 100% RH) : 0÷ 20mA, 4÷ 20mA, 0÷ 1V, 0÷ 5V, 1÷ 5V, 0÷ 10V
    Range: 0 ÷ 100% RH
    Resolution 0.1 % RH
    Accuracy ± 2.5% RH @ 23° C
    Analog Outputs
    • Number 5, selectable between all available output quantities
    • Range 0÷ 20mA, 4÷ 20mA, 0÷ 1V, 0÷ 5V, 1÷ 5V, 0÷ 10V
    • Resolution 14 bit max
    Power supply
    • Range 12 ÷ 30 VDC
    • Power